Preparing Electronic Materials for RoHS Testing

Analytical testing of electronic materials requires representative preparation of often heterogeneous components. The preparation should support the substance measurements selected by the laboratory without introducing relevant contamination.

Contact Glen Mills for information or a quote
RETSCH RS 200 Vibratory Disc Mill for fine grinding hard, brittle samples; placing a grinding set on the rail of the instrument
RS 200 Vibratory Disc Mill
01

Does a milled electronic sample demonstrate RoHS compliance?

No. Compliance depends on the current applicable requirements and the complete testing and assessment process, not the milling operation alone.

In July 2006, the EU Directive Restriction of Hazardous Substances (RoHS) for restricted use of certain hazardous substances such as lead, mercury, cadmium and brominated flameproofing agents in electrical and electronic products became effective. An update in March 2015 (2015/863) added four new substances, mainly present in plastic parts, namely Bis(2-ethylhexyl)phthalate (DEHP); Benzylbutylphthalate (BBP); Dibutylphthalate (DBP); Diisobutylphthalate (DIBP). In July 2019, an amended RoHS Directive with new prohibitions on the use of substances came into force.

All EU countries must adopt the directive through national legislation which will not only affect EU members but also countries which export their products to Europe. In other parts of the world, like California and China, the directive has also been transformed into national regulations. In any case, the manufacturers of electric appliances as well as recyclers and disposers of electronic scrap need to carry out analytical examinations to prove that the RoHS directive is being observed.

02

Sample preparation of electronic scrap

The industries involved look for testing protocols for these regulated substances and while in most cases approved analytical methods are available, one major issue is often omitted: correct sample preparation. The affected electrotechnical products come in many forms and sizes from relatively simple devices such as toys all the way through to complex instruments and computers each of which contains many different sub-assemblies. Hence, it can be a challenge to produce a representative, neutral-to-analysis laboratory sample for subsequent analysis.

The primary target of investigation are electronic assemblies such as PCBs and individual components, plastic housing parts, mains cables and wiring. In PVC insulation, the detection of cadmium, lead and DEHP is required. It is important to remove any iron and steel components before starting the sample preparation process. These need to be evaluated separately.

To carry out reproducible analyses, and achieve representative results, a completely homogeneous sample with a fineness below 500 µm is required. With such a heterogeneous mixture of materials made from metals, glass, ceramics and plastics, this is no easy task. In addition, electronic components may have different breaking properties which make the selection of a suitable laboratory mill more difficult.

For the initial size reduction of this type of material, a RETSCH Cutting Mill such as the SM 300 (left) is recommended. This high-performance mill has proved to be particularly suitable for the size reduction of a heterogeneous mixture of hard and ductile as well as soft-elastic materials and larger initial feed size up to 80 mm. The complete laboratory sample, which can be as much as 30 liters when a cyclone system is used, is processed to a fineness of less than 7 mm in a first step. The SM 300 is available in a heavy-metal-free version which makes it especially suitable for sample materials which are analyzed for traces of hexavalent chrome. For the typical material mixes found in electronic scrap, the use of the 6-disc rotor is beneficial as it acts more like a shredder than the

parallel section rotor which acts like an axe. Therefore, also electric cables or tough materials are easily comminuted.

Application examples Cutting Mill SM 300

Electronic and polymer sample pre-crushing examples
SampleAccessoriesStarting size and quantitySpeed and grinding timeFinal fineness D90
Electronic scrap• universal hopper • 6-disc rotor with cutting tips tungsten carbide • bottom sieve 4 mm • cyclone170 mm 100 g3000 rpm 30 s< 5 mm
Electronic scrap with larger metal parts• universal hopper • 6-disc rotor with cutting tips tungsten carbide • bottom sieve 6 mm50 mm 50 g1500 rpm 10 s< 7 mm
Keyboard• universal hopper • 6-disc rotor with cutting tips tungsten carbide • bottom sieve 4 mm • cyclone80 mm 1 piece1500 rpm 7 min< 5 mm

With the help of a sample divider (such as RETSCH’s PT

100, left, or, for sample amounts up to 60 l, the PT 600) a

representative

sub-sample

is obtained

for

further

pulverization to analytical fineness.

The required grind size of 500 µm and less can be easily achieved with the Ultra- Centrifugal Mill ZM 200 (right). The two-stage rotor/ring sieve system is extremely gentle on the material. Equipped with

tungsten-reinforced

grinding tools, wear through abrasion is not an issue with the ZM 200, and the risk of introducing impurities during milling can be avoided. This extremely efficient mill is characterized by very short grinding times and easy and convenient operation. Other RETSCH mills which are suitable for fine grinding of this type of sample material are the Vibratory Disc Mills RS 200 or RS 300 (pulverization of sample amounts of up to 2 l in one go) and the Mixer Mill MM 400 (for small sample volumes of up to 2 x 20 ml).

The new regulations also affect plastic softeners. RETSCH’s

CryoMill

(left)

facilitates

the

pulverization of elastic, ductile materials, which would only be deformed at room temperature, by automatic embrittlement of the sample with liquid nitrogen during grinding. In the ZM 200, the sample material can easily be embrittled with liquid nitrogen or dry ice prior to grinding. The dry ice may be fed into the mill during grinding which

has the additional benefit of chilling the sample while it is pulverized. This not only improves the breaking properties and avoids frictional heat but also ensures that volatile fractions are not adversely affected. The ZM 200 processes prechilled materials without any problems.

Application examples for pulverization of electronic scrap and plastic materials

Electronic and polymer sample fine grinding examples
SampleMill / accessoriesStarting size and quantitySpeed and grinding timeFinal fineness D90
Electronic scrap• Vibratory Disc Mill RS 200 • grinding set 250 ml hardened steel5 mm; 70 g1500 U/min; 4 min< 500 µm
Electronic scrap with larger metal parts• Vibratory Disc Mill RS 200 • grinding set 50 ml hardened steel7 mm; 50 g1500 U/min; 1 min< 500 µm
Keyboard• Ultra-Centrifugal Mill ZM 200 • push-fit rotor 12 teeth, stainless steel • ring sieve 0.5 mm; cyclone • cryokit and LN25 mm; 1 piece18000 U/min; 15 min< 550 µm
Rubber• CryoMill • grinding jar 50 ml stainless steel • 1 x 25 mm ball stainless steel10 mm; 4 g8 min pre-cooling, 4 x (2 min grinding at 30 Hz and 30 s intermediate cooling at 5 Hz)< 200 µm
Capacitor• CryoMill • Grinding jar 35 ml stainless steel • 1 x 20 mm ball stainless steel5 mm; 3 g10 min precooling, 5 x (2 min grinding at 30 Hz and 30 s intermediate cooling at 5 Hz)< 150 µm

Plan the configuration for your material.

Define the component or material being assessed before reducing it. Separation of different materials, tool contact and the final analytical portion can affect interpretation. Discuss preparation with the laboratory performing the test; historic equipment applications do not establish current regulatory scope or compliance.

Preparing Electronic Materials for RoHS Testing: application checklist
ConsiderationInformation to provide
Test materialComponent or defined homogeneous material
Target analysisSubstances and required method
Preparation controlSeparation and tool contributions

How should the application examples be used?

The results describe the materials and configurations discussed here. Confirm equipment options, contact materials and operating conditions for your own sample before adopting a method. A demonstrated endpoint does not guarantee the same outcome for a different material.

Discuss material testing with Glen Mills or request further information or a quote.